Green Liquid Chromatography for Simple Applications

Authors

  • J. Šesták Department of Chemistry and Technology of Environmental Protection, Faculty of Chemistry, University of Technology, Brno | Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Brno
  • V. Kahle Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Brno

Keywords:

green chemistry, green analytical chemistry, green liquid chromatography, nanocolumns

Abstract

A simple laboratory setup is described for performing green liquid chromatography in a nanocolumn with a great potential to minimize solvent consumption and waste production as required in green chemistry. The system consists of a high-pressure syringe fixed in vertical position and with weight. This setup acts as a constant-pressure pump. Samples are injected by short switch of a two-way six-port injection valve equipped with an external sample loop. Isocratic separation of EPA 8330 standards was performed. The relative standard deviations of retention times below 1 % demonstrate a constant mobile phase flow for several hours.

Published

2015-06-15

How to Cite

Šesták, J., & Kahle, V. (2015). Green Liquid Chromatography for Simple Applications. Chemické Listy, 109(6), 470–474. Retrieved from http://ww.w.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/364

Issue

Section

Articles